Cannon-instrument CMRV-4500 Manual de usuario Pagina 46

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 128
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 45
40
CANNON
®
Mini-Rotary Viscometer CMRV-4500 Instruction & Operation Manual
Version 1.1e—October, 2011 • CANNON
®
Instrument Company
2139 High Tech Road • State College, PA • 16803 • USA
16. When acceptable values have been obtained, click Abort Tempera-
ture Control. Then click Save Data to save the new calibration
constants.
Do not omit this step or new calibration data will be lost.
17. Close the Calibrate Cell Constants window and then click Main/
Save Instrument. The Save Instrument window will appear.
18. In the Save As box, input a name for the new instrument configura-
tion (which includes the newly-calculated calibration constants).
Then click OK to save the configuration.
Printing calibration data
CANNON Instrument Company recommends generating a hard copy
record of calibration data after each calibration session. To print the
current calibration constants, click Configure/Print Instrument Set-
tings.
Vista de pagina 45
1 2 ... 41 42 43 44 45 46 47 48 49 50 51 ... 127 128

Comentarios a estos manuales

Sin comentarios