The Instrument Settings window
4. Use the ID field to input instrument identification information using
up to 30 alphanumeric characters.
The remaining fields in the Instrument Settings window are non-
editable (information is obtained via serial communication with the
CMRV instrument). The Model: field will indicate the model of your
instrument. The S/N: field indicates the serial number from the label
on the rear service panel. The Firmware: field indicates the version
for the current instrument firmware. The Port: field indicates the
current communications port for the RS-232 cable connecting your
computer to the CMRV-4500 instrument.
5. When you have entered all settings, click OK.
Checking initial calibration
After you have completed the CMRV instrument calibration for tempera-
ture and cell constants (see following chapters), follow the procedure
below to verify the initial calibration settings.
1. From the VISCPRO primary window, click Configure/MRV/View
Cell Constants. The View Cell Constants window will open.
2. Ensure that Calibration data, including constants and viscosity, is
available for each cell. Then click OK to close the View Cell Con-
stants window.