60
Service menu
To access the service menu items, click Service from the VISCPRO
®
primary display and select the desired instrument.
The Monitor window permits the user to view current operational values
for the miniPV Series instrument. These settings may be helpful to
CANNON
®
technical personnel when troubleshooting difficulties with
the miniPV Series.
Select Wash viscometer tube to open the Wash viscometer tube
window.
Click Begin Wash to initiate the wash cycle. The Status: indicator will
display and update current actions.
Bypass .... option
To begin washing an evacuated sample cup, click the option box to make
certain that the Bypass initial cup evacuation option is selected.
CAUTION
Do not use this option with a filled cup or the cup will overflow.
When the Wash cycle has been completed, click Done to exit the Wash
viscometer tube window.
NOTES
The Wash operation cannot be executed if the tube is currently running a
sample. To terminate the Wash cycle at any time, click Abort Wash.
CAUTION
The Wash viscometer tube function overrides automatic miniPV Series
software operations. Verify that the tubes are clean and dry before initiat-
ing computer-controlled sample testing.