Cannon-instrument miniPV-X Manual de usuario Pagina 59

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 130
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 58
53
CANNON
®
miniPV-Series Automatic Viscometer with VISCPRO
®
Instruction & Operation Manual
Version 1.a— January, 2011; CANNON
®
Instrument Company
2139 High Tech Road • State College, PA 16803 • USA
Selecting sample run required and maximum determination
settings
Select the number of Required determinations by clicking on the
spin
controls (values from 1 to 4 are acceptable). In the same manner, select
the Maximum number of determinations which can be run for the tube
samples (values from 1 to 20 are acceptable).
NOTES
VISCPRO determines the sample flow time by averaging the values of the
sample runs which correspond to the chosen criteria.
If the Always perform maximum determinations radio button is selected,
the last flow times will be used for the viscosity determination.
When you have completed the configuration and saved the current set-
tings, you may wish to permanently store the settings using the Save
Instrument and Tray Settings.
Tray Settings: Wash
Use the Tray Settings: Wash window to determine wash settings. For
multi-tube units, different wash settings may be selected for each tube
individually.
The Tray Settings: Wash window
Solvent rinse settings
The solvent rinse settings for Solvent A and (if applicable) Solvent B are
determined using the Tray Settings: Wash window. Select a value from
0 to 18 cycles for this option (the “0” setting turns the solvent option off).
These settings determine the number of rinse cycles performed by the
Vista de pagina 58
1 2 ... 54 55 56 57 58 59 60 61 62 63 64 ... 129 130

Comentarios a estos manuales

Sin comentarios