miniPV-Series Automatic Viscometer with VISCPRO
®
Instruction & Operation Manual
Version 1.a— January, 2011; CANNON
®
Instrument Company
2139 High Tech Road • State College, PA 16803 • USA
56
The Tray Settings: Advanced window
Maximum bulb 1 fill time(s):
Adjust this setting to select the desired maximum fill time for bulb 1.
Acceptable values are between 20 and 400 seconds.
Min. bulb 1 selection time(s):
Adjust this setting to select the desired minimum time permitted for
selection of bulb 1 or bulb 2 for sample testing. Acceptable values are
between 0 and 20 seconds. Decreasing the setting will increase the
probability that the sample will be tested in bulb 1.
Maximum bulb empty time(s):
Adjust this setting to select the maximum amount of time for the oil to
drop between the target tube sensor and the bottom tube sensor during a
sample determination before the sample test is aborted. Acceptable
values are between 90 and 600 seconds.