Calibrating the TESC System CANNON®
During testing, the Brookfield® viscometer measures and records the temperature of the TESC
base unit via an RTD probe located in the thermal core behind the test chamber. Adjusting
the offset of the Brookfield® viscometer only enables the recording of an accurate sample
temperature. It does not affect the operation of the TESC System.
Table 12Brookfield® offset example
Brookfield®
Display
Offset Watlow® Display
Sample
Temperature
41 °C
0
40 °C
40 °C
40 °C
+1 °C
40 °C
40 °C
Watlow® controller calibration
The Watlow® controller must be calibrated before adjusting the temperature probe offset of
the Brookfield® viscometer.
Table 13Tools and materials required
Description
Description
Assembled TESC System with mounted
Brookfield® DV2T viscometer
20 mL of an appropriate reference sample
25 mm × 150 mm rimless, disposable test
tube
Brookfield® #4B2 insulated spindle
Sample preparation materials
RS-485 to USB cable
USB flash drive with program files
Install Watlow® configurator program
A copy of the Watlow® EZ-ZONE Configurator 5.0 program is provided with the TESC
software package. This program must be loaded onto a computer running a Windows®
operating system prior to calibrating or operating the TESC System.
1. To load the Watlow® EZ-ZONE Configurator 5.0 program, run USB flash drive> Watlow >
EZCv5.exe.
2. Follow the instructions provided by the installation wizard. For support and additional
information, visit http://www.watlow.com.