SimpleVIS
®
Viscometer
Reference
Guide
P82.0090
Vista de pagina 0
1 2 3 4 5 6 ... 15 16

Indice de contenidos

Pagina 1 - Reference

SimpleVIS®ViscometerReferenceGuideP82.0090

Pagina 2

Emptying the Waste Bottle ReceiverCheck the volume of the Waste Bottle frequently to avoid spillage/overflow.Disconnect the Waste Bottle from the cap

Pagina 3 - Viscometer

either of the two following calibration intervals. Performance Based Method Determine need to calibrate by performing weekly check-runs of certified v

Pagina 4

with the reference with the desired tolerance.Changing the temperature calibration by more than 0.05°C will affect the instrument viscosity calibratio

Pagina 5 - Running a Test

Unpack your SimpleVIS® ViscometerRemove all packing materials from the components. The SimpleVIS® Viscometer should be placed on a stable laboratory t

Pagina 6

14vacuum line by approximately 2 inches.4. Connect the other ends of both pieces of tubing as indicated in the photo image.Select Target Temperature1

Pagina 7

Waste BottleMedia CapP82.007515Printer Cable(optional)P82.0065Printer Power Supply(optional)Printer with Power Supply(optional)P82.0076PipettorP82.005

Pagina 8

2139 High Tech Road • State College • PA • 16803814-353-8000 • 800-676-6232 • Fax [email protected] • www.cannoninstrument.comsi

Pagina 9

Sample/Solvent PortRun Status Indicator LampsTouch Screen DisplayRS-232 SerialPort Vacuum LinePower ConnectorWaste LineExhaust MufflerDust CoverWaste

Pagina 10 - Maintenance & Calibration

SAFETY CAUTIONS Please observe the following safety procedures and notices for proper operation of the SimpleVIS® Viscometer.• Make sure that you rea

Pagina 11

The ~MAINS symbol indicates instructions or connections for the AC power supply. The AC Power input must match the electrical specifications listed

Pagina 12

GETTING STARTEDPress the on switch on the AC/DC adapter. The Warm up Screen will display a “PLEASE WAIT” message. During this time, the instrument wil

Pagina 13 - Basic Information

Injecting a sample1. Using the supplied piston operated pipettor, insert the tip into a vial of test sample and pull the measured amount into the pip

Pagina 14

Print the results (only if optional printer is installed)1. Press the Print button on the display screen to printout the test results to an attached s

Pagina 15 - Components for

Dry CyclePress the Dry button on the display wash screen to start the dry run cycle. The display will show the time remaining in the dry cycle.When th

Pagina 16

The SimpleVIS® Viscometer was designed to require minimal maintenance through simplicity of design and built-in electronic diagnostics. All new instru